GDCR3000B Fa'afuainumera Fa'atekonolosi Fa'atekonolosi Su'ega
GDCR3000B Earth Resistance/Soil Resistivity Tester e faʻapitoa faʻapitoa mo le fuaina o le teteʻe o le eleele, le malosi o le eleele, le eletise o le eleele ma le AC voltage.Ua fa'aaogaina le fa'atekonolosi fa'akomepiuta ma masini fa'atekonolosi, e fua ai le tete'e o le eleele i le 4-pole, 3-pole po'o le 2-pole.FFT (Fourier transform) tekinolosi, AFC (otometi fa'atonutonu fa'asolosolo) o lo'o maua fo'i le tekonolosi, fa'atasi ai ma le lelei o le fa'alavelave fa'alavelave ma le malosi e fa'afetaui ai le si'osi'omaga, e fa'amautinoa ai le sa'o maualuga, mautu maualuga ma le fa'amaoni.E faʻaaogaina lautele mo le fuaina o le eletise, fesoʻotaʻiga, meteorology, suauʻu fanua, fausiaina, faʻamalamalamaga puipuiga, mea tau eletise.
GDCR3000B o loʻo aofia ai masini talimalo, mataʻituina polokalama, faʻataʻitaʻiga uaea, tootoo eleele fesoasoani, uaea fesoʻotaʻiga ma isi.O le tele LCD faʻaaliga o masini talimalo o loʻo i ai le lanu moana lanu moana ma le kalafi pa e faʻaalia ai e mafai ona iloa manino.I le taimi lava e tasi e mafai ona teuina 300 seti o faamatalaga, iloa fesili talafaasolopito ma luga o le initoneti galuega mataituina taimi moni e ala i le mataituina polokalama, faʻaalia malosi, faʻailoga faʻailoga, faʻatasi ai ma galuega o le faitau, toe iloilo, faʻasaoina, lipoti ma le lolomiina o faʻamatalaga faʻasolopito.
1. Tulaga masani ma Tulaga Galuega
Tulaga masani | Tulaga Galuega | Fa'amatalaga | |
vevela o le siosiomaga | 23℃±1℃ | -10℃~40℃ | ---- |
Si'osi'omaga Susū | 40%~60% | <80% | ---- |
Galulue Voltage | 9V±0.1V | 9V±1.5V | ---- |
Auxiliary Earth Resistance | <100Ω | <30kΩ | rC、rP |
Fa'alavelave fa'alavelave | ---- | <20V | ---- |
Fa'alavelave i le taimi nei | ---- | <2A | ---- |
Electrode mamao ole su'ega R | a>5d | a>5d | ---- |
Electrode mamao oρsuega | a>20h | a>20h | ---- |
2. Fa'amatalaga Lautele
Galuega | 2-pou/3-pou/4-pouFuaina o le tetee i le eleele,tetee i le eleele,eletise eleele, AC voltagefua. |
Paoa sapalai | DC 9V(Zn-Mn maa mago R14S 1.5V 6 vaega,Fa'aauautaimi tutumau300itula) |
Avanoa | Tete'e o le Lalolagi: 0.00Ω~30.00kΩ |
Tete'e o le eleele: 0.00Ωm~9000kΩm | |
Eleele eletise: 0V-600V | |
Ituaiga fua | 4-pole,3-pole, 2-pole fua o le tete'e eleele. |
Auala ole Fua | Tete'e o le eleele: Fa'avasegaina pole suiga auala.Su'ega ole Max.20mA Tete'e o le eleele: 4 pole method (Wenner Method). Volosa o le lalolagi: Fa'asa'o masani[P(S)-ES taulaga] |
Su'ega Fa'atele | 128Hz/111Hz/105Hz/94Hz(AFC) |
Su'ega Fa'ata'ita'i Pu'upu'u i le taimi nei | AC 20mA maualuga |
Tatala Su'ega Su'ega Voltage | AC 40V maualuga |
Su'ega Galulue Galu | Galu o le Sina |
Va'aiga o Electrodes Mamao | E mafai ona seti1m-100m |
Suiga Tulaga | Tete'e o le Lalolagi: 0.00Ω-30.00kΩ Suia ta'avaleing. |
Tete'e o le eleele: 0.00Ωm-9000kΩm Kilia ta'avaleing. | |
Molī tua | moli lanu moana i tua |
Fa'aali Fa'ata'ita'iga | 4 fa'aaliga LCD numera, moli lanu moana i tua |
Malamalama Fa'ailoga | Emo fa'ailoga LED ile taimi ole su'ega, LCD faitau i lalo fa'aaliga |
Fuaina LCD | 128mm*75mm |
Fa'aaliga LCD | 124mm*67mm |
Fua | 215mm * 190mm * 95mm |
Ta'ita'i Su'ega Tulaga | Aofa'i 4 ta'ita'i: 1* Mumu20m, 1* Uuliuli20m, 1* Samasama10m, 1* Meamata10m |
Su'ega Faigofie | 2 ta'ita'i: 1* Mumu 1.6m,1*uliuli 1.6m |
Auxiliary Ground Electrodes | 4 vaega:φ10mm×150mm |
Fua Taimi | I le eletise eletise: e tusa ma le 3 taimi / sec |
Tete'e o le eleele, tetee i le eleele: e tusa ma le 5 sekone/taimi | |
Taimi Fua | E sili atu i le 5000 taimi (su'ega pu'upu'u, su'ega 1 taimi, taofi le 30 sekone ona su'e ai lea) |
Malosi o le Circuit | Su'e i lalo ole AC 600V |
Taulaga feso'ota'iga | USB interface, taulotoina faamatalaga e mafai ona uploaded i PC e teu ma lolomi. |
Feso'ota'iga Ta'ita'i | USB,1.5m * 1 fasi |
Teuina o Faamatalaga | 300 vaega,“MEM”fa'ailoga le teuina,“ATUNUA”aikona e taʻu mai ai ua tumu le teuina |
Taofi Faamatalaga | Fa'amatalaga taofi galuega: "TASI" fa'aaliga ata |
Fa'amatalaga Faitau | Fa'amatalaga faitau galuega: "FAITAU" fa'aaliga ata |
Fa'aaliga Ova | Ole tele ole fua ole fua ole galuega fa'amama: “OL” fa'aaliga fa'ailoga |
Su'ega Fa'alavelave | Faailoa fa'ailoga fa'alavelave fa'afuase'i, wmatuamoavoltage fa'alavelavesili atu5V,e iai "PISO” faailoilo faailo |
Su'ega eleele fesoasoani | 0.00KΩ-30kΩ(100R+rC<50kΩ, 100R+rP<50kΩ) |
Galuega Fa'ailoga | A sili atu le tau o le fuaina o le fa'ailo, e iaileofa'ailo |
Malosi o le ma'a | Pe a fa'aitiitia le malosi o le maa i le tusa ma le 7.5V, maa eletise maualalo aikona" "o le a faaalia, faamanatu itotogi |
Fa'aaogāga Malosiaga | Faatali mai:e uiga i20mA(Pepe le moli tua) |
Molī tua: e tusa ma le 45mA(Tape moli tua 25mA) | |
Fua:e tusa ma le 100mA (Tape moli tua) | |
mamafa | Aofa'i mamafa ma ato: 4.5kg |
Tagata su'e: 950g(e aofia ai maa) | |
Su'ega Uea: 1560g | |
Auxiliary earthing rod: 935g (4PCS) | |
Galulue Vevela & Susū | -10℃-40 ℃, lalo ole 80%rh |
Teuga vevela & susū | -20℃-60℃, lalo ole 70%rh |
Puipuiga ovata | Fuaina o le malosi o le eleele: i le va o feso'ota'iga ta'itasi oC(H)-E、P(S)-ES, AC 280V/3 sekone |
Insulation Resistance | Ua uma20MΩ(i le va o le matagaluega ma le pa puipui e 500V) |
Malosiaga Malolo | AC 3700V/rms (I le va o le taamilosaga ma le pa puipui) |
Fa'ailoga Electromagnetic | IEC61326(EMC) |
Ituaiga Puipuiga | IEC61010-1(CATIII 300V, CAT IV 150V, tulaga filogia2); IEC61010-031; IEC61557-1(Tete'e o le Lalolagi); IEC61557-5(Tete'e o le eleele); JJG 366-2004 |
3. Fa'atonu sese ma le fa'atinoga i lalo o tulaga masani
Aitema | Va'aiga fua | Sa'o | I'ugafono |
Tete'e o le eleele(R) | 0.00Ω~30.00Ω | ±2%rdg±3dgt | 0.01Ω |
30.0Ω~300.0Ω | ±2%rdg±3dgt | 0.1Ω | |
300Ω~3000Ω | ±2%rdg±3dgt | 1Ω | |
3.00kΩ~30.00kΩ | ±4%rdg±3dgt | 10Ω | |
Tete'e o le eleele(ρ) | 0.00Ωm~99.99Ωm | E tusa ai ma le sa'o o le R(ρ=2πaR a:1m-100m; π=3.14) | 0.01Ωm |
100.0Ωm~999.9Ωm | 0.1Ωm | ||
1000Ωm~9999Ωm | 1Ωm | ||
10.00kΩm~99.99kΩm | 10Ωm | ||
100.0kΩm~999.9kΩm | 100Ωm | ||
1000kΩm~9000kΩm | 1kΩm | ||
Malosiaga eleele | AC 0.0~600V | ±2%rdg±3dgt | 0.1V |
Fa'amatalaga:1. fa'aopoopo mea sese: ≤±3%rdg±5dgt i lalo o le rC max po'o le rP max.(rC max: 4kΩ+100R<50kΩ, rP max: 4kΩ+100R<50kΩ)
2. fa'aopoopo mea sese: ≤±5%rdg±5dgt i lalo ole 5V fa'alavelave voltage