GD3128 Fa'asologa Insulation Resistance Tester
GD3128 series Insulation Resistance Tester o loʻo i ai le suʻesuʻega atoatoa o le tele o faʻamaufaʻailoga faʻafefeteina ma le lelei tele o le faʻalavelave faʻalavelave, lea e mafai ona faʻaaogaina e faʻataʻitaʻi ai le faʻaogaina o le malosi tele eletise eletise eletise ma laina faʻasalalau i le siosiomaga malosi e pei o le substation.O le meafaigaluega e faʻaaogaina le lanu paʻi lanu moni, o faʻamaumauga uma o suʻega ma le malosi o le maa o loʻo faʻaalia i luga o le mata lava e tasi i le tepa, ma o le tuʻufaʻatasia o le ki ma le ki e faigofie tele ai le gaioiga.E le manaʻomia e le tagata faʻaoga le taulotoina o le auala faʻaoga.Kiliki "Fesoasoani" i luga o le lau e oso i luga le auala faʻagaioiga, lea e mafai ona taʻitaʻia lelei le tagata faʻaoga e faʻaogaina le meafaigaluega.O le meafaigaluega o loʻo i ai le galuega o le faʻasaʻo vave, lea e mafai ona tuʻuina atu le voltage o le mea faʻataʻitaʻi faʻafuaseʻi e aunoa ma se faʻasalalauga faʻaopoopo.O le mataʻituina o le eletise o le meafaigaluega e mafai ona otometi lava ona mataʻituina le voltage ola o le mea faʻataʻitaʻi.Afai o le voltage e sili atu i le 36V, o le a otometi lava ona faʻasaina le suʻega, lea e mafai ona puipuia lelei le meafaigaluega ma faʻalapotopotoga.
O le meafaifa'aili e fa'aaogaina le pusa maa lithium e mafai ona fa'aaogaina.Ole taimi ole su'ega ole meafaifa'aili e otometi lava ona fa'amaumau taimi ole su'ega ma teu fa'ai'uga o su'ega ile aso ma le taimi.O le pa'i pa'i e mafai ona faigofie ona maua fa'amaumauga fa'asolopito fa'amaumau.E mafai e le USB fa'aesea atoatoa ona tu'uina atu ma le saogalemu fa'amatalaga su'ega i le PC.O le meafaigaluega ua faʻapipiʻiina i le laina faʻataʻitaʻi faʻatasi ai ma le faʻaogaina faʻalua ma le talipupuni, ma o le paʻu eletise maualuga ua faʻapipiʻiina i le kilipa alligator sui ma le matau, lea e mafai ona faʻafetaui i nofoaga eseese o suʻega.O le meafaigaluega e faʻaaogaina se fausaga faʻapipiʻi faʻapipiʻi lua, ma o le puipuiga o le pusa i fafo o le IP65, lea e mafai ona puipuia ai le osofaʻiga o le susu ma le pefu i le taimi o felauaiga ma le teuina, ma puipuia lelei le meafaigaluega mai le faʻalavelave.
GD3128 faʻasologa o loʻo i ai le lautele lautele, maualuga maualuga, faʻaogaina faigofie, malosi ma le tumau, saʻo ma le faʻamaoni, faʻamautu le faʻatinoga, ma le malosi malosi e tetee atu i le faʻalavelave.O le meafaigaluega o loʻo i ai le suʻega faʻafefe (IR);su'ega fa'asinomaga polarization (PI);su'ega ole fua ole absorption (DAR);su'ega fa'asolosolo (RAMP);su'ega faamama (FR, 10S, 20S, 30S, 40S fa filifiliga);suega eletise (V);su'ega o le malosi (uF);su'ega o iai nei (nA) ma isi galuega.O le meafaigaluega e fetaui lelei mo le suʻega faʻafefeteina o masini eletise, laina faʻasalalau ma isi masini tetele i lalo o le siʻosiʻomaga o le gafatia tele, maualuga maualuga ma le malosi o le faʻaogaina.O se meafaigaluega e masani ona faʻaaogaina ma manaʻomia mo le eletise eletise, fesoʻotaʻiga, meteorology, potu komepiuta, suauʻu fanua, faʻapipiʻi eletise eletise ma le tausiga, ma le eletise tuʻuina atu matagaluega o alamanuia ma mining pisinisi.
●Ole maualuga ole va'ava'a ole 7mA.
●Su'ega fa'agata (IR), su'ega fa'asinomaga polarization (PI), su'ega fua faatatau absorption (DAR).
●Fa'ata'ita'iga fa'ata'ita'i (RAMP), fa'ata'ita'iga fa'amama (FR), su'ega capacitance (uF)
●Voltage mata'ituina galuega, mata'ituina otometi le voltage molia o le mea ua tofotofoina.
●Fa'atino taimi su'ega, fa'amaumau taimi o su'ega.
●Otometi fa'ate'a galuega, otometi ma vave fa'asa'oloto le moliaga o le su'ega mea pe a uma le su'ega, e aunoa ma se fa'alavelave fa'ate'a fafo.
●Otometi tapuni galuega.
●Lau lanu pa'i lapoa.
●Fausia atigi fa'alua-lau, malosi ma umi, puipuiga grade IP65.
●Ma'a lithium e mafai ona fa'aaogaina maualuga.
●Galuega Teuina, e mafai ona otometi ona teuina 1000 vaega o faʻamaumauga faʻataʻitaʻi taimi faʻatasi ma le aso ma le taimi o suʻega ma le taimi o suʻega.
●Faʻapipiʻi galuega, faʻapipiʻiina i le USB interface, faʻamaumauga faʻamaumauga e mafai ona faʻapipiʻiina i le komepiuta mo suʻesuʻega faʻamaumauga o faʻamaumauga.
Fa'ata'ita'iga o oloa | Malosi Fa'atatau | Fa'a'ese'ese Fa'asagaga Va'aiga | Va'a Puupuu |
GD3128A | 250V,500V,1kV,2.5kV,5kV | 0.50MΩ~10.0TΩ | ≥7mA |
GD3128B | 250V,500V,1kV,2.5kV,5kV,10kV | 0.50MΩ~20.0TΩ | ≥7mA |
GD3128C | 250V,500V,1kV,2.5kV,5kV,10kV,15kV | 0.50MΩ~30.0TΩ | ≥7mA |
Insulation Resistance Ranga ma le sa'o
Malosi Fa'atatau | Insulation Resistance ita | Sa'o | Fa'amatalaga |
250V | 0.50MΩ~50.0GΩ | ±5%rdg±3dgt |
|
50.0GΩ~500GΩ | ±20%rdg±3dgt |
| |
500V | 1.00MΩ~100GΩ | ±5%rdg±3dgt |
|
100GΩ~1.00TΩ | ±20%rdg±3dgt |
| |
1000V | 1.00MΩ~200GΩ | ±5%rdg±3dgt |
|
200GΩ~2.00TΩ | ±20%rdg±3dgt |
| |
2500V | 1.00MΩ~500GΩ | ±5%rdg±3dgt |
|
500GΩ~3.00TΩ | ±20%rdg±3dgt |
| |
5000V | 1.00MΩ~1.00TΩ | ±5%rdg±3dgt |
|
1.00TΩ~10.0TΩ | ±20%rdg±3dgt |
| |
10KV | 1.00MΩ~2.00TΩ | ±5%rdg±3dgt |
|
2.00TΩ~20.0TΩ | ±20%rdg±3dgt |
| |
15KV | 1.00MΩ~3.00TΩ | ±5%rdg±3dgt |
|
3.00TΩ~30.0TΩ | ±20%rdg±3dgt |
1 TΩ (Tera ohm)=1000GΩ=1012Ω
1 GΩ (Giga ohm)=1000MΩ=109Ω
1 MΩ (Mega ohm)=1000KΩ=106Ω
Galuega | Fuaina o le fa'agata (IR);fua fa'asinomaga polarization (PI);fua faatatau o le mitiia (DAR);faiga fa'ata'ita'i (RAMP);faamamagaluega faatino(10S, 20S, 30S, 40S);vfua fa'atatau (V);gapatitancefua(uF);fua i le taimi nei(nA) |
Paoa sapalai | 6.2Ah maa lithium toe faʻaaogaina |
Malosi Fa'atatau | GD3128A: 250V, 500V, 1kV, 2.5kV, 5kV GD3128B: 250V, 500V, 1kV, 2.5kV, 5kV, 10kV GD3128C: 250V, 500V, 1kV, 2.5kV, 5kV, 10kV, 15kV |
Sa'o lelei o Voltage Output | (5%~10%) ±10V |
Su'ega Su'ega Fa'asa'o | GD3128: 0.50MΩ~10.0TΩ GD3128B: 0.50MΩ~20.0TΩ GD3128C: 0.50MΩ~30.0TΩ |
Fa'aola Pu'upu'u Fa'aola | ≥7mA |
Su'ega Fa'asinomaga Polarization | Ioe |
Su'ega Fa'atatau ole Absorption | Ioe |
Faiga Su'ega Ramp | Fa'ata'ita'i i la'asaga o le 10% o le voli fa'atonu se'ia o'o i le vili fa'atonu. |
Filifili Su'ega Faiga | 10S, 20S, 30S, 40S fa filifiliga. |
Su'ega Volta | Avanoa: AC/DC: 0V~1000V; Sa'o: ±5%rdg±3V |
Su'ega Gapa | Avanoa: 10nF~25uF; Sa'o: ±10%rdg±10nF |
Su'ega o lo'o iai nei | Avanoa: 0.01nA~7mA; Sa'o: ±5%rdg±0.5nA |
Mataitu Malosi | Mataitu le voltage o lesu'egad mea faitino, ma mataʻituina le tuʻuina atu o le tuʻuina atu pe a uma le suʻega, faʻasaina le suʻega pe a sili atu le eletise i le 36V, puipuia meafaigaluega ma le faʻaogaina. |
Taimi Su'ega | Otometi fa'amaumau taimi su'ega, timingva'aiga: 0s~9999s |
Galuega Teuina | Otometi teuina faʻamatalaga suʻega ma le aso ma le taimi o suʻega, aofaʻi o 1000 vaega |
Tu'u atu Galuega | Tu'u le fa'amaumauga o lo'o teuina i le komepiuta e ala i le uaea feso'ota'iga USB. |
Fa'aaliga Malosiaga ma'a | Faatasi ai ma le faʻaalia o le malosi o le maa, pe a maualalo le voltage o le maa o le a faamanatu i aitotogile maa. |
Tapuni otometi | A maeʻa le 15 minute amata o le a tapuni otometi e aunoa ma se gaioiga. |
Fua | 280mm×260mm×160mm |
mamafa | 4900g (e aofia ai le maa) |
Uea Su'e | Uea mumu lanu mūmū su'esu'e volt 1pc (maa kilipi alligator ma matau), uaea su'ega lanu meamata 1pc, uaea su'ega lanu uliuli 1pc |
Puipuiga Tulaga | Tapuni le pusa IP65, tatala le pusa IP40 |
Siosiomaga Galuega | -20℃~50℃; 80%rh |
Siosiomaga Faleoloa | -25℃~65℃; 80%rh |
Insulation Resistance | 50MΩ(1000V) (i le va o le taamilosaga su'ega ma atigi) |
Taofi Malolo | AC 3kV50Hz 1min(i le va o le taamilosaga su'ega ma atigi) |
Saogalemu | IEC61010-1, IEC61326-1 |